page 134 008-0608-00
14.5.2
DIAGNOSTICS
Sub-Menu
DAC FULL SCALE
Menu Item
When this menu item is selected, the Analog Output of the channel is forced to its full-
scale output, then
DAC UPDATED
is displayed. This is useful when calibrating or trimming
the readout connected to the Analog Output.
DAC ZERO SCALE
Menu Item
When this menu item is selected, the Analog Output of the channel is forced to its
zero-scale output, then
DAC UPDATED
is displayed. This is useful when calibrating or trim-
ming the readout connected to the Analog Output.