Quick Reference Guide
Automatic Test and Calibration Station
Vista de página 0
1 2 3 4 5 6 ... 18 19

Resumo do Conteúdo

Página 1 - Quick Reference Guide

iERP: 126102D5618/7 [English]© BW Technologies 2008. All rights reserved.Quick Reference GuideAutomatic Test and Calibration Station

Página 2

MicroDock IIQuick Reference Guide8The base station pushbuttons are not labelled. The base station is operated by pressing the C pushbutton that is loc

Página 3

9MicroDock IIBump Test5. Press C exit to return to normal operation.For more information on setting up inlets, gas types, and con-centration levels, r

Página 4 - Display Elements Pushbuttons

MicroDock IIQuick Reference Guide10A calibration adjusts the sensor sensitivity to ensure an accu-rate response to gas.To calibrate a detector, comple

Página 5 - User Options Menu

11MicroDock IIAutomatic Datalog Download (GasAlertMicroClip and GasAlertMax XT docking module only)Successful Data Transfer: The PASS LED lights green

Página 6 - Installation

MicroDock IIQuick Reference Guide12• GasAlertMicroClip• GasAlertMax XTThe detector can be configured prior to performing a bump test or calibration to

Página 7

13MicroDock IIMaintenance4. Allow the battery to obtain a full charge (2 to 4 hours, depending upon how many docking modules are con-nected to this st

Página 8 - Inserting the Detector

MicroDock IIQuick Reference Guide14SpecificationsThe MicroDock II base station and docking modules are for indoor use only.Instrument dimensions: (w x

Página 9 - Using the Base Station

15MicroDock IICharger SpecificationsCharger SpecificationsSize: 8.6 x 8.2 x 7.8 cm (3.4 x 3.2 x 3.1 in.)Weight: 97 g (3.4 oz.) per modelCharger system

Página 10 - MicroDock II

MicroDock IIQuick Reference Guide16

Página 11 - Bump Test

iERP: 126102D5618/7 [English]© BW Technologies 2008. All rights reserved.Quick Reference GuideAutomatic Test and Calibration Station

Página 12 - = calibration has failed

Limited Warranty and Limitation LiabilityBW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal

Página 13 - Eventlogging

1MicroDock IIIntroductiona WarningTo ensure personal safety, read Safety Information - Read First before using the MicroDock II base station.The Micro

Página 14 - (Optional)

MicroDock IIQuick Reference Guide2Display Elements PushbuttonsItem FunctionAC powerBatteries fully chargedBatteries half-chargedBatteries at low level

Página 15 - Maintenance

3MicroDock IIUser Options MenuUser Options MenuDescription DisplayTime/Date modifies the time and date of the base station.Inlet Setup enters the gas

Página 16 - Specifications

MicroDock IIQuick Reference Guide4InstallationItem Description1 Inlet filter assembly2 C-cell batteries (4)3 Battery cover4 Philips pan head retaining

Página 17 - Charger Specifications

5MicroDock IIInstallationa WarningThe atmosphere must be free of background gas. Do not use the base station in a hazardous area.All required National

Página 18

MicroDock IIQuick Reference Guide6Inserting the Detectora CautionInfrared or intense ambient light (sun or halogen) may interfere with the base statio

Página 19

7MicroDock IIUsing the Base Station• When the detector has been inserted correctly, the RUN LEDs light yellow, Unit Inserted displays on the base stat